Title: Potentiometric performance of 2-aminothiophenol based dipodal ionophore as a silver sensing material
Authors: Mittal, S. K.
Ashok Kumar, A.
Kaur, S.
Kumar, S.
Keywords: 2-Aminothiophenol
Selectivity
Dipodal
Indicator electrode
Liquid membrane electrode
Potentiometric titration
Issue Date: 2007
Abstract: A new sensing material has been used for fabricating a liquid membrane electrode for silver determination. The electrode has been systematically characterized by the electrode response, measuring range (1 × 10-4 to 1 × 10-1 M), detection limit (6 × 10-5 M), pH range (3.3-8.0) and selectivity coefficients for the closely related metal ions. The electrode shows a good response to silver ions in the presence of mercury ions. The electrode works well in mixed solvents media also. The electrode is used extensively as an indicator electrode for the determination of Ag+ in various synthetic samples and also used for the determination of SCN-, CN-, S2- and I-. © 2006 Elsevier B.V. All rights reserved. © 2008 Elsevier B.V., All rights reserved.